IMPORTANT!
Operator safety
Do not place your hand in between the crosshead and the compression
plate unless the measurement device is on MODE 0! To operate the clamps,
use manual switches located in the clamp body (operator panel does not
work in 0 mode).
In case the machine must be stopped immediately, press the red button
on the control panel. To release the button, rotate it clockwise.
Upper safetylimit is 10 mm for force controlled pulling experiment (e.g.
paper creep). It takes more than 10 mm for clamps to stop moving.
General user rules
Log out after your experiments for all machines.
If somebody is logged in, you are not allowed kick him out. For example, if
you are trying to contact the computer remotely, click ”No”, when remote
desktop asks: ”User XYZ is logged in and his session will be closed. Do
you want to proceed?”
1
Vista de pagina 0
1 2 3 4 5

Indice de contenidos

Pagina 1 - General user rules

IMPORTANT!Operator safety• Do not place your hand in between the crosshead and the compressionplate unless the measurement device is on MODE 0! To ope

Pagina 2 - 1 Foreword

1 ForewordThis is a list of instructions to bring the Instron measurement device and nec-essary camera gear on-line and to perform wood fatigue tests.

Pagina 3 - 9 After experiments

• After all looks clear, switch the machine again on MODE II.In some wood compression experiments, you can also put around the sampleblack cardboard p

Pagina 4 - 10 Instron initialization

10 Instron initializationFirst, before anything else is done with the measurement device, it must beinitialized properly. In the case of compression t

Pagina 5

11 FAQ11.1 How can I calibrate the force and position meters?• Open the Instron panel and choose either Position or Load tab, dependingon which you wa